1066 (2-wire)

Model 1066 Single Channel pH, ORP, Contacting and Toroidal Conductivity 2-Wire Transmitter & Analyzer

  • Chemically & Mechanically Resilient Polycarbonate NEMA 4X / CSA 4 IP66 Enclosure – Standard 1/2 DIN Cutout
  • 2-Wire Loop-Powered 24VDC Power Operation with scalable 4-20mA analog current loop output standard

    • Second 4-20mA loop-powered output available to send process temperature reading in addition to measured analytical liquid parameter

  • Automatic Temperature Compensation via 100 or 1000 Ohm Platinum Temperature Compensation Element (Available in Standard and ACCU-TEMP fast response configurations)
  • HART 7 is Standard and Foundation FieldBus available as an optional digital output
  • Large LCD Display with touch membrane keypad – Menu Driven Interface and Programming
  • Automatic Temperature Compensation from 0 to 150 °C (32 to 302 °F) for pH/ORP and 0 to 200 °C (32 to 392 °F) for Conductivity
  • Area Classifications:

    • Intrinsically Safe (with appropriate safety barrier) for:
      • CSA/UL Class I, II, III, Div. 1 Groups A-G T4 Tamb = -20°C to 65°C

      • ALTEX Baseefa04ATEX0195X EEx ia IIC T4 Tamb = -20°C to 65°C
      • IECEx BAS 11.90098X EEx ia IIC T4 Tamb = -20°C to 65°C

    • Non-Incendive:

      • CSA/UL Class I, Div. 2, Groups A-D Dust Ignition Proof Class II & III, Div. 1, Groups E-G NEMA 4X Enclosure T4 Tamb = -20°C to 65°

1066 Product Specifications

1066 Operation Manual

Measurement Input Measurement Range Outputs Calibration Points Compatible Sensor(s) Special Features
pH/ORP Single or Dual Channel
– pH/ORP
– 0 to 14 pH  (standard)
– Fully Scalable from 1 to 13 pH units
– Analog  0-20 mA or 4-20 mA output for pH/ORP/ISE or temperature for each input channel
– Optional HART or ProfiBUS Digital Outputs
– 2 point auto buffer recognition for pH for slope determination
– 1 point user defined pH standardize calibration to correct for offset (drift)
– Any Suitable ASTI pH/ORP  Sensor with 100 or 1000 Ohm Platinum TC
– Any Suitable ASTI pH/ORP  Sensor with 100 or 1000 Ohm Platinum TC & 1056 compatible preamplifier
– Excellent option for severe service pH & ORP measurement in areas with flammable gas and corrosive environments.
Contacting Conductivity Single or Dual Channel
– Conductivity Cell
– Cell from 0.01 to 10.0, user selectable
– Ranges from 0-200 microSiemens (0.01/cm) to 0-200 milliSiemens (10.0/cm) as mates with cell
– Analog  0-20 mA or 4-20 mA output for Conductivity or temperature for each input channel
– Optional HART or ProfiBUS Digital Outputs
– Zero Calibration (Capitance)
– Cell Constant calibration tofFind exact effective (apparent) cell constant in standard solution or process media
– Any Suitable Contacting Conductivity  Sensor with 1000 Ohm Platinum TC – Support for displaying in concentration units of acids, bases and electrolytes as well as salinity
– Special ultrapure water temperature compensation and support for display in resistivity units
Toroidal Conductivity (Contactless Inductive) Single or Dual Channel
– Toroidal Conductivity Sensor
– Range from 0.050 to 2,000 milliSiemens (2 Siemens) – Analog  0-20 mA or 4-20 mA output for Conductivity or temperature for each input channel
– Optional HART or ProfiBUS Digital Outputs
– Zero Calibration (Capitance)
– Cell Constant calibration tofFind exact effective (apparent) cell constant in standard solution or process media
– Any Suitable Toroidal Conductivity  Sensor with 20/20 Windings and 1000 Ohm Platinum TC – Support for displaying in concentration units of acids, bases and electrolytes as well as salinity
– Excellent choice for strong acid, strong base and strong electrolyte solutions at elevated temperatures

https://www.astisensor.com/1066.php